Intracranial volume using the reference, automatic, and corrected-automatic methods from 19 subjects

Reference (mL) Automatic (mL) Corrected-Automatic Obs1 (mL) Corrected-Automatic Obs2 (mL)
Mean 1504 ± 169 1509 ± 172 1504 ± 174 1505 ± 173
Median 1545 1566 1550 1553
Range 1180–1761 1194–1776 1187–1776 1189–1777