Skip to main content
. 2013 Nov;34(11):2192–2198. doi: 10.3174/ajnr.A3531

Table 4:

Comparison of arterial damage by the device samples according to the presence or absence of a clot and the number of retrievalsa

With Clot (10 Samples) Without Clot (10 Samples) P Valueb Two Retrievals (10 Samples) Three Retrievals (10 Samples) P Valuec
Mural thrombus (%) 0.5 ± 1.6 10.1 ± 19.3 .21 3.2 ± 10.3 7.3 ± 17.6 .31
Endothelial denudation (%) 81.8 ± 18.9 63.8 ± 35.8 .25 68.9 ± 28.8 76.6 ± 30.0 .41
Intimal layer edema (%) 35 ± 33.4 99.1 ± 57 .009 84.1 ± 62.8 50 ± 45.6 .24
IEL fractured (%) 10 ± 17.5 37.5 ± 61.5 .18 40 ± 61.5 7.5 ± 12.1 .27
Medial layer edema (%) 45 ± 43.8 59.1 ± 26.5 .44 56.6 ± 37 47.5 ± 36.2 .58
a

Data are means +/− standard deviations.

b

Mann-Whitney test for samples with versus without clots.

c

Mann-Whitney test for 2 versus 3 retrievals.