. 2014 Dec;35(12):2237–2242. doi: 10.3174/ajnr.A4056

Table 3:

Diagnostic confidence per reader for evaluation of ventricular size

100% Dose
25% Dose
10% Dose
J30 J30 H30 J30 H30
Reader 1 5.0 ± 0.0 5.0 ± 0.0 5.0 ± 0.2 4.7 ± 0.5 4.6 ± 0.5
Reader 2 5.0 ± 0.2 4.8 ± 0.5 4.6 ± 0.6 3.9 ± 0.9 3.3 ± 1.2
Reader 3 5.0 ± 0.0 5.0 ± 0.0 5.0 ± 0.0 4.9 ± 0.5 4.7 ± 0.6

Note:—5-point scale: 1 indicates nondiagnostic, cannot identify or rule out ventricular enlargement and shunt tube tip location; 2, will potentially miss mild ventricular enlargement and shunt tube tip location; 3, will probably not miss ventricular enlargement and shunt tube tip location; 4, most likely will identify all abnormalities with respect to ventricular enlargement and shunt tube tip; 5, can detect ventricular enlargement and shunt tube tip without diagnostic compromise.