Published in final edited form as: Med Phys. 2021 Feb 6;48(3):1211–1225. doi: 10.1002/mp.14681

Fig. 12.

Fig. 12.

Projection images of the Catphan spatial resolution phantom (reconstructed images in Fig. 12) and pelvis sized Gammex electron density phantom (reconstructed images in Fig. 9). Only a zoomed in section of the Gammex phantom is shown, and hypointense object is the cortical bone insert. Row (a), raw projection: intensity reduction in 2D grid’s shadows is clearly visible. Row (b), flat-field corrected projection: Since residual scatter was not corrected, septal shadows are overcompensated, and appear as hyperintense lines. Row (c), scatter map: Residual scatter intensity as measured by the GSS method. Row (d), GSS Method: Residual scatter was subtracted from the raw projection, followed by flat-field correction.