Skip to main content
. Author manuscript; available in PMC: 2022 Mar 1.
Published in final edited form as: Med Phys. 2021 Feb 6;48(3):1211–1225. doi: 10.1002/mp.14681

TABLE III.

HU nonuniformity in water-equivalent sections

Maximum HU difference and standard deviation among nine ROIs

Head Pelvis
Max difference Standard deviation Max difference Standard deviation
2D grid only 39 HU ±12.3 HU 65 HU ±21.7 HU
2D grid + GSS 32 HU ±10.3 HU 30 HU ±10.2 HU
MDCT 12 HU ±4.7 HU 21 HU ± 6.4 HU