Table 2:

Average systematic error (M), standard deviation of the systematic error (Σ), and standard deviation of the random error (σ) of the x, y, and z isocenter shifts during daily setup of 10 patients (253 adaptive fractions on the MR-linac).

Axis M (mm) Σ (mm) σ (mm)
x (left/right) −0.3 3.3 1.7
y (sup/inf) 0.7 2.6 2.9
z (ant/post) 0.1 1.4 2.0