Skip to main content
. 2020 Sep 7;25(4):2203–2211. doi: 10.1007/s00784-020-03534-8

Table 2.

Multiple linear regression analysis with maximum implant failure strength (N) as dependent variable and implantoplasty, implant type/design, diameter, and material as predictors

Parameter Coefficient 95% Confidence interval p value
Lower Upper
Implantoplasty No 0.0 < 0.001
Yes − 82.90 − 97.40 − 68.40
Implant type/design Bone level 0.0 < 0.001
Tissue level − 139.96 − 154.46 − 125.46
Implant diameter Narrow 0.0 < 0.001
Regular 347.28 332.74 361.82
Implant material Ti 0.0 0.473
TiZr 5.27 − 9.23 19.77

Ti (titanium) and TiZr (titanium zirconium) alloy. Bold values indicate statistical significance