. 2014 Jul;35(7):1288–1292. doi: 10.3174/ajnr.A3851

Table 1:

Quantitative assessment of metal artifacts before and after reduction

Objects HU Range
P Value Noise
P Value
Before After % Change Before After % Change
ROI0 (n = 28) 247.2 ± 37.5 247.3 ± 37.3 0.04 NS 36.3 ± 4.8 36.2 ± 4.7 0.2 NS
Coiling ± stenting (n = 20) 2005.6 ± 804.0 282.6 ± 55.8 85.9 <.001 251.7 ± 94.4 42.9 ± 10.3 82.9 <.001
    <25 cm (n = 6) 1613.7 ± 709.7 296 ± 44.7 81.7 .006 198.4 ± 83.8 42.8 ± 7.3 78.4 .007
    25–100 cm (n = 7) 1954.9 ± 634.2 257.6 ± 23.5 86.8 <.001 238.6 ± 49.3 38.9 ± 5.6 83.7 <.001
    >100 cm (n = 7) 2392.3 ± 945.6 296 ± 80.7 87.6 .001 310.3 ± 113.5 46.9 ± 15.0 84.9 .001
Liquid embolizer (n = 6) 1377.3 ± 646.6 428.5 ± 259.0 68.9 .019 206.0 ± 86.3 56.9 ± 21.6 72.3 .007
Shunt (n = 2) 1226.5 ± 259.5 368.0 ± 76.4 70.9 NS 109.8 ± 2.6 44.0 ± 0.7 60 .013
Overall (ROI1) (n = 28) 1815.3 ± 793.7 319.9 ± 136.6 82.3 <.001 231.7 ± 95.9 45.9 ± 14.0 80.2 <.001

Note:—NS indicates no significance.