Skip to main content
. 2021 Mar 3;18(5):2482. doi: 10.3390/ijerph18052482

Table 2.

High-concern organisms identified during the second phase of microbiological surveillance.

Type of Endoscope
(Number of Samples)
Microorganism Swab from Biopsy Channel Swab from Suction Channel Swab from Air/Water Channel Swab from Elevator Channel Liquid from Internal Channels
High-Concern Organisms
Duodenoscope
(10)
S. maltophilia 0 0 0 0 1/10 (10%)
K. pneumoniae
K. oxytoca
0 0 0 0 4/10 (40%)
(1/4 KPC-K. oxytoca)
Total viable Count (quantity range and % NC samples) - - - - 50–>300 CFU
2/10 (20%)
Gastroscope
(1)
P. fluorescens 0 0 0 0 1/1 (100%)
Total viable count (quantity range and % NC samples) - - - - >300 CFU
1/1 (100%)
Colonoscope
(6)
P. aeruginosa
P. fluorescens
0 0 2/6(33.3%) 0 3/6(50%)
Total viable count (quantity range and % NC samples) - - - - 20–>300 CFU
4/6 (66%)

NC: non-compliant (TVC > 20 CFU/channel or >100 CFU/channels, according to ESGE/ESGENA or CDC guidelines, respectively).