Table 1. Optimized MRM Transitions, Retention Times, Method Limits of Detection and Quantitation, and Linear Ranges for EDTs and DHDTs.
analyte | Q1 (Da) | Q3 (Da) | tR (min) | LOD (pM) | LOQ (pM) | linear range (nM) |
---|---|---|---|---|---|---|
16,17-DHDT | 365.5 | 235.0 | 6.44 | 50 | 100 | 0.125–200 |
13,14-DHDT | 365.5 | 195.1 | 6.58 | 50 | 100 | 0.125–200 |
10,11-DHDT | 365.5 | 154.9 | 6.67 | 10 | 50 | 0.125–200 |
7,8-DHDT | 365.5 | 96.9 | 6.78 | 100 | 500 | 0.625–200 |
16,17-EDT | 347.2 | 246.9 | 8.25 | 5 | 10 | 0.0125–200 |
13,14-EDT | 347.2 | 194.9 | 8.44 | 5 | 10 | 0.0125–200 |
10,11-EDT | 347.2 | 182.9 | 8.50 | 10 | 50 | 0.125–200 |
7,8-EDT | 347.2 | 134.9 | 8.71 | 500 | 1000 | 1.25–200 |