Table 2.
Quantification error results for spheres simulated in a cylindrical phantom with and without partial volume corrections (PVCs)
| Object | Quantification error (%) | |||
|---|---|---|---|---|
| Without PVC | With PVC | Without PVC | With PVC | |
| Sphere | Sphere-CF | Sphere-CF-RC | Cylinder-CF | Cylinder-CF-RC |
| 15.6 ml | − 4.56 | 3.32 | − 20.08 | 3.28 |
| 24.4 ml | − 3.40 | 0.19 | − 19.11 | 0.19 |
| 33.5 ml | − 2.05 | − 0.43 | − 17.98 | − 0.43 |
| 65.4 ml | 1.60 | 0.83 | − 14.93 | 0.86 |
| Average | − 2.10 ± 2.67 | 0.98 ± 1.64 | − 18.03 ± 2.23 | 0.98 ± 1.62 |
CF Calibration factor; PVC Partial volume correction, RC Recovery coefficient