Table 3.
Quantification error results for spheres simulated in the torso phantom with and without partial volume corrections (PVCs)
| Object | Quantification error (%) | |||
|---|---|---|---|---|
| Without PVC | With PVC | Without PVC | With PVC | |
| Sphere | Sphere-CF | Sphere-CF-RC | Cylinder-CF | Cylinder-CF-RC |
| 15.6 ml | − 4.92 | 2.92 | − 20.39 | 2.99 |
| 24.4 ml | − 5.54 | − 2.03 | − 20.90 | − 2.03 |
| 33.5 ml | − 4.84 | − 3.26 | − 20.32 | − 3.26 |
| 65.4 ml | − 2.70 | − 3.44 | − 18.53 | − 3.41 |
| Average | − 4.50 ± 1.24 | − 1.45 ± 2.98 | − 20.04 ± 1.04 | − 1.43 ± 3.01 |
CF Calibration factor; PVC Partial volume correction, RC Recovery coefficient