Table 4.

Radiation dose from [18F]FDG and 99mTc-nanocolloid as received by operating theatre per procedure for each staff groups

Variable Category Total Operating surgeon Assisting surgeon Scrub nurse Anaesthesiologist Anaesthetic technician Missing data
Number of staff members n 393 66 63 132 66 66
Measured EPD dose [µSv] n 388 66 62 128 66 66 5 (1.3%)
Mean (SD) 38.2 (38.1) 61.8 (45.5) 54.8 (48.9) 33.4 (32.5) 26.4 (22.4) 20.0 (20.7)
Median [Q1–Q3] 27.0 [11.0–56.0] 56.5 [38.0–74.0] 44.5 [27.5–66.0] 16.0 [4.0–59.0] 22.0 [16.0–28.8] 14.0 [10.0–20.0]
Min–max 0.0–335.0 0.0–335.0 3.0–262.0 0.0–140.0 3.0–135.0 3.0–113.0
Finger badge dose received "Left" [µSv] n 104 32 29 23 20 0 289 (73.5%)
Mean (SD) 57.6 (26.8) 72.9 (21.2) 53.8 (17.9) 64.6 (33.0) 31.2 (15.4)
Median [Q1–Q3] 58.0 [38.0–75.0] 75.0 [58.0–94.3] 48.3 [38.0–66.0] 64.0 [30.8–91.4] 20.0 [16.7–45.0]
Min–max 16.7–104.0 43.3–104.0 33.3–82.9 16.7–100.0 16.7–64.0
Finger badge dose received "Right" [µSv] n 99 33 33 27 6 0 295 (75.1%)
Mean (SD) 62.2 (31.3) 61.9 (30.4) 63.9 (25.8) 51.5 (13.0) 101.7 (75.1)
Median [Q1–Q3] 60.0 [53.0–70.0] 60.0 [56.0–60.0] 70.0 [53.0–70.0] 60.0 [46.5–60.0] 99.2 [33.3–165.0]
Min–max 29.0–220.0 33.3–220.0 29.0–145.0 29.0–60.0 33.3–180.0