Skip to main content
. 2006 May;27(5):989–994.

Table 3:

Mean operator exposure rate measurements during kyphoplasty (KP), vertebroplasty with a CDS (VP-CDS), and vertebroplasty with syringes (VP-S)

KP VP-CDS VP-S
Needle/device placement 0.81 ± 0.76 1.05 ± 0.89 0.32 ± 0.33
    95% CI 0.6–1.0 0.9–1.3 0.2–0.5
Cement delivery 1.74 ± 3.87 2.88 ± 2.71 0.21 ± 0.35
    95% CI 0.9–2.6 2.3–3.5 0.05–0.4
Mean total exposure rate 1.14 ± 1.63 1.9 ± 1.53 0.3 ± 0.27
    95% CI 0.8–1.5 1.6–2.2 0.2–0.4

Note:— Measurements (μSv/min) include means ± SD.