Skip to main content
. 2006 Jan;27(1):60–66.

Table 1:

Influence of kVp, CBR, slip composition, and orientation on artifact volume

Ratio
KVp*
    90 vs 120, High mode 1.3
    120 vs 140, High mode 1.2
    90 vs 120, Standard mode 1.4
    120 vs 140, Standard mode 1.2
Conebeam reconstruction vs linear interpolation* 1.1
Cobalt-containing alloy vs Titanium clip#
    High mode 1.3
    Standard mode 1.5
In plane vs perpendicular 1.4

Note:—CBR indicates conebeam reconstruction.

*

for titanium- and cobalt-containing alloy clip.

#

for all kVp/mAs combinations.

for both clips in high and standard mode.