Table 1:

Conditions used to expose the programmable valves to the 3T MR system

Test Pressure Setting Orientation Arrow Direction
1 0 Sagittal Up
2 0 Sagittal Down
3 0 Axial Up
4 0 Axial Down
5 0 Coronal In
6 0 Coronal Out
7 8 Sagittal Up
8 8 Sagittal Down
9 8 Axial Up
10 8 Axial Down
11 8 Coronal In
12 8 Coronal Out
13 20 Sagittal Up
14 20 Sagittal Down
15 20 Axial Up
16 20 Axial Down
17 20 Coronal In
18 20 Coronal Out
19 5 Simulated patient In
20 5 Simulated patient Out

Note:—All conditions were in-out × 5. There were no changes under any conditions.