Table 2:

Sequences run on patient 2

Acquisition Time (min:sec) TR (ms) TE (ms) Echo Train Resolution IPAT AccelerationFactor Comments
T1 SAG and Cor FSE 2:18 475 16 3 256 × 256 2 Nondiagnostic due to motion
T2 Ax FSE 2:22 5890 104 21 256 × 205 2 Nondiagnostic due to motion
T1 GRE, 3 planes 0:32 102 4.8 NA 256 × 156 2 Nondiagnostic due to motion
T1 FSE with navigator, 3 planes 4:00–6:00 429 10 3 256 × 205 2 Diagnostic
T2 Ax FSE 4:30 5890 104 21 256 × 205 2 Diagnostic

Note:—IPAT indicates integrated parallel acquisition technique; SAG, sagittal; Cor, coronal; FSE, fast spin-echo; Ax, axial; GRE, gradient-echo; NA, not applicable. Both the TR and the acquisition time will be a variable of the frequency of patient motion. They should be adjusted inversely with concatenations to keep the effective TR short for T1 sequences.