Skip to main content
. 2007 Apr;28(4):628–634.

Table 3:

Possible causes for disagreement between TOF MRA and DSA in 16 patients

Location N (%) Size (mean, mm) Overestimation (Size) Underestimation (Size)
ICA 40 (21.3) 10.4 Vessel overlap (9 mm) Pulsation-induced artifacts (10 mm)
AcomA 65 (34.6) 5.7 Vessel overlap (8 mm) Susceptibility at bone/air interface (2.5 mm)
Vessel overlap and small size (2 mm, 3 mm)
Partially thrombosed aneurysm (10 mm)
Pulsation-induced artifacts (2 mm, 12 mm)
ACA 7 (3.7) 9.1
MCA 26 (13.8) 7.1 Vessel overlap (11 mm)
BA 23 (12.2) 7.6 Pulsation-induced artifacts (10 mm) Slow flow and small size (2 mm)
Saturation effects and small size (3 mm)
PcomA 23 (12.2) 7.3 Vessel overlap (10 mm)
Signal void due to slow flow (5 mm)
Susceptibility at bone/air interface (3 mm)
PICA 3 (1.7) 6.7
VA 1 (0.5) 4.0
Total 188 (100) 7.5

Note:—MRA indicates magnetic resonance angiography; DSA, digital subtraction angiography; ICA, internal carotid artery; AcomA, anterior communicating artery; ACA, anterior cerebral artery; MCA, middle cerebral artery; BA, basilar artery; PcomA, posterior communicating artery; PICA, posterior inferior cerebellar artery; VA, vertebral artery.