. 2021 Feb 17;13(2):e13393. doi: 10.7759/cureus.13393

Table 2. Monte Carlo (MC) simulations to determine average Root Mean Square Errors (RMS-e) were computed across the entire volume enclosed by the N-localizer system for various heights subtended by the Brown-Roberts-Wells (BRW) localizer frame (LF). RMS-e decreases with increasing the width/height ratio. The BRW LF (standard) is the commercial version, whereas the other versions are modified in only the vertical dimension.

Millimeter (mm)

Root Mean Square Error (RMS-e)

BRW, Radionics CRW Stereotactic System, Integra LifeSciences Corporation, Plainsboro, New Jersey

Localizer Width (mm) Height (mm) Width/Height Average RMS-e Whole space (9 rods)
BRW LF (standard) 140 189 0.741 0.512 (+/- 0.014)
BRW LF (modified) 140 160 0.875 0.451 (+/- 0.011)
BRW LF (modified) 140 140 1 0.410 (+/- 0.010)
BRW LF (modified) 140 120 1.167 0.372 (+/- 0.008)