Skip to main content
. 2007 Aug;28(7):1384–1387. doi: 10.3174/ajnr.A0557

Table 3:

Aneurysm volume and relevant packing results of 35 aneurysms treated with helical GDC 18 coils compared with 35 volume-matched controls treated with complex Trufill/Orbit coils

GDC 18 Trufill/Orbit P value (t test)
Number of aneurysms 35 35
Aneurysm volume .98
    Mean 297 mm3 295 mm3
    Median 400 mm3 406.5 mm3
    Range 26–1450 mm3 27–1504 mm3
    SD 327 mm3 383 mm3
Largest aneurysm diameter
    Mean 10.3 mm 9.6 mm
    Median 10 mm 10 mm
    Range 4–22 mm 4–17 mm
Packing .15
    Mean 23.1% 25.1%
    Median 24% 25%
    Range 8%–33% 11%–36%
    SD 5.98% 5.67%
Σ Aneurysm volumes 14 002 mm3 14 229 mm3
Σ Inserted coil length 2794 cm 4446 cm
Coil length per mm3 aneurysm volume 0.20 cm 0.31 cm
Mean coil length per aneurysm per mm3 0.224 cm 0.339 cm <.0001
Total number of coils 157 238
Mean number of coils per aneurysm 4.5 6.8 .0024