Published in final edited form as: Int J Environ Anal Chem. 2015 Dec 21;96(1):15–37. doi: 10.1080/03067319.2015.1114104

Table 3.

Accuracy, precision and uncertainty of the XOS prototype instrument based on analysis of certified reference materials.

CRM Matrix Element Certified (mg/kg) ±U XOS prototype (mg/kg) %RSD %Bias n Variability
BCR 627 Tuna Fish Tissue As 4.8 0.3 5 0 4 5 Within-day

DORM 2 Dogfish Muscle As 18 1.1 17 9 −7 10 Within-day
Cr 34.7 5.5 23 29 −33
Cu 2.34 0.16 2 0 −15
Fe 142 10 102 19 −28
Mn 3.66 0.34 ND - -
Hg 4.64 0.26 3 15 −40
Ni 19.4 3.1 11 22 −44
Se 1.4 0.09 2 23 29
Zn 25.6 2.3 25 28 −2

SRM 612 Glass Matrix Ba* 41 75.5 30 84 5 Within-day
Co* 35.5 1.2 43.6 5 23
Cu* 37.7 0.09 51 5 35
Au* 5 ND - -
Fe 51 2 36.4 8 −29
Mn* 39.6 0.8 24.4 8 −38
Ni 38.8 0.2 50.6 6 30
Ag 22 0.3 18.2 18 −17
Sr 78.4 0.2 62.4 2 −20
Ti* 50.1 0.8 ND - -
Pb 38.57 0.2 35.4 3 −8
K* 64 267 71 317
Rb 31.4 0.4 29.4 3 −6

SRM 2976 Mussel Tissue As 13.3 1.8 13 4 −4 5 Between-day
Pb 1.19 0.18 2 39 77
Br* 329 289 3 −12
Ca* 7600 6055 9 −20
Cl* 57000 52184 5 −8
Fe 171 4.9 157 9 −8
K* 9700 7610 7 −22
Mn* 33 28 10 −16
P* 8300 7974 9 −4
Rb* 4.14 5 16 10
S* 19000 14332 2 −25
Se 1.8 0.15 1 59 −31
Sr* 93 66 3 −29
Zn 137 13 137 6 <1
*

reference or informational value; ±U is the expanded uncertainty from the CRM certificate of analysis; PE=polyethylene