Table 1:

Summary of MR Imaging Parameters

3D T1w TSE 3D TOF
Low-resolution High-resolution

Field of viewa 230x230x210mm3 230x210x180mm3 105x180x190mm3
Voxela (Acq./Rec.) 1.5x1.5x1.5mm3/0.72x0.72x0.75mm3 0.5x0.5x0.5mm3/0.24x0.24x0.25mm3 1.0x0.5x0.5mm3/0.5x0.25x0.25mm3
Flip angle 90° 18°
Echo train lengthb 5 40 30 40 N/A
VFA designc SNR prior PSF prior SNR prior N/A
Echo spacing/timed 3ms 5ms 3.45ms
Repetition time 800ms 20ms
Driven-equilibrium Anti-drive N/A
Fat suppression Spectral adiabatic inversion recovery No
Orientation Sagittal Axial
Acceleration CS-SENSE x5 CS-SENSE x6 SENSE(RL) x2.2
Halfscan(RL) 0.75
Scan time 7:39 0:58 8:50 6:38 6:37
a

The imaging field of view and voxel size is represented in a format of FHxAPxRL.

b

The echo train length for turbo spin echo (TSE) scan only considers the number of acquired echoes, and the skipped start-up echoes at the beginning are not counted.

c

The SNR and PSF prior used in this VFA design category correspond to SNR and PSF priority VFA design approach, respectively. The PSF priority approach is the default VFA design on our scanner that tries to achieve an ideal/flat PSF (the first 11 echoes) given the minimum flip angle (50 degree) and start to decay once the calculated flip angle over a prescribed threshold (120 degree).

d

Echo spacing and echo time correspond to TSE and time-of-flight (TOF) scan respectively. TE was not a fixed value for TSE scans since different number of skipped echoes were investigated.