Skip to main content
. 2020 Oct 2;31(4):2349–2356. doi: 10.1007/s00330-020-07315-5

Table 3.

Summary of test statistics for intraoperative CBCT assessment of screws

Parameter Value CI (95%)
Sensitivity 90% (9/10) 55–100%
Specificity 98% (244/250) 95–99%
Diagnostic accuracy 97% (253/260) 95–99%
PPV 60% (9/15) 32–84%
NPV 100% (244/245) 98–100%

CI confidence interval, NPV negative predictive value, PPV positive predictive value