Table 2.

Comparison of hemodynamic measurements between the invasive method and ClearSight.

Variable Invasive method ClearSight Percentage error Bias LOA
PR 73.68 ± 12.37 73.88 ± 12.27 11.65 0.21 −8.22 to 8.63
Blood pressure; mmHg
 Systolic 109.93 ± 17.18 104.76 ± 18.29 31.01 −5.18 −37.81 to 27.45
 Diastolic 63.45 ± 9.50 68.61 ± 10.78 27.14 5.16 −12.41 to 22.72
 Mean 79.95 ± 11.46 80.99 ± 12.47 25.23 1.05 −18.85 to 20.95

Values are presented as mean ± standard deviation.

LOA = limits of agreement, PR = pulse rate.