Table 2.

Ranges of calibration, limits of detection (LOD), and limits of quantification (LOQ) expressed as concentration in matrix, recovery (%), and expanded measurement uncertainty (%) for all elements.

Element (γ) Calibration
range (γ)
LOD (γ) LOQ (γ) Recovery
(%)
Measurement uncertainty
U (%)
Al (mg L−1) 0.1–2.0 0.0004 0.0015 93 5
As (μg L−1) 15–300 9 28 97 11
B (mg L−1) 0.25–5.0 0.001 0.002 94 12
Ba (mg L−1) 0.05–1.00 0.00001 0.00005 99 5
Ca (mg L−1) 5.0–100.0 0.004 0.014 101 15
Cd (μg L−1) 1–20 0.3 0.9 101 4
Co (μg L−1) 0.5–10.0 0.4 1.2 106 11
Cr (μg L−1) 0.5–10.0 0.3 1.2 106 16
Cu (mg L−1) 0.05–1.00 0.0004 0.0013 98 5
Fe (mg L−1) 0.5–10.0 0.003 0.009 101 7
K (mg L−1) 100–2,000 0.05 0.16 106 8
Li (μg L−1) 1–20 0.004 0.014 93 7
Mg (mg L−1) 5–100 0.002 0.005 99 7
Mn (mg L−1) 0.25–5.00 0.00004 0.00014 101 6
Mo (μg L−1) 0.5–10 0.1 0.2 107 33
Na (mg L−1) 1.0–20.0 0.00 0.01 98 7
Pb (μg L−1) 15–300 5 16 101 5
Rb (mg L−1) 0.25–5.00 0.0003 0.0009 96 9
Sn (μg L−1) 0.05–1.00 0.004 0.012 99 5
Sr (mg L−1) 0.05–1.00 0.000004 0.000014 100 7
V (μg L−1) 5–100 0.3 1.1 107 8
Zn (mg L−1) 0.25–5.00 0.0005 0.0018 101 7