TABLE 3.

Calculated doses compared to measured dose for different material assignments, anthropomorphic phantom with largest clinically realistic cement dimensions.

Energy/mode Field size (cm) ABS Average %diff (Eclipse calc vs Measured)
PVC PTFE Titanium CT HU
Acuros XB AAA Acuros XB AAA Acuros XB AAA
6X 3x3 7.9 6.5 6.0 4.1 9.5 7.4
5x5 7.5 5.9 5.8 3.9 8.8 6.5
10x10 6.4 4.8 4.8 2.6 8.6 6.6
VMAT 3.3 7.0
6SRS 3x3 7.8 6.4 5.9 4.0 9.7 7.9
5x5 7.5 5.9 5.8 3.8 9.4 7.4
10x10 6.2 4.5 4.6 3.3 9.2 7.4
VMAT 3.3 8.0
6FFF 3x3 7.5 6.9 5.4 4.2 10.3 7.6
5x5 7.0 6.6 5.2 4.3 9.8 6.8
10x10 5.6 5.3 3.9 3.2 9.6 6.9
10X 3x3 6.9 5.7 5.2 4.3 7.5 4.5
5x5 6.0 4.0 4.2 2.4 7.5 4.4
10x10 5.7 3.7 3.9 2.2 7.5 4.7
10FFF 3x3 6.6 5.1 4.8 3.4 8.0 5.0
5x5 6.1 4.2 4.3 2.5 8.1 5.0
10x10 5.3 3.4 3.5 1.7 8.0 5.2
VMAT 2.9 6.3
Average % diff all modes: 6.7 5.2 4.6 4.0 8.8 6.2

PVC, polyvinyl chloride, PTFE, polytetrafluoroethylene; CT, computed tomography (CT); HU, Hounsfield units.