Published in final edited form as: AJR Am J Roentgenol. 2018 May 7;211(1):W33–W41. doi: 10.2214/AJR.17.18702

TABLE 2:

Multiparametric MRI Acquisition Parameters at 3 T

Parameter T2-Weighted DWIa High-b-Value DWIb DCE-MRIc
FOV (mm) 140 × 140 140 × 140 140 × 140 262 × 262
Acquisition matrix 304 × 234 112 × 109 76 × 78 188 × 96
TR (ms) 4434 4986 6987 3.7
TE (ms) 120 54 52 2.3
Flip angle (°) 90 90 90 8.5
Section thickness (mm), no gap 3 3 3 3
Image reconstruction matrix 512 × 512 256 × 256 256 × 256 256 × 256
Reconstruction voxel imaging resolution (mm/pixel) 0.27 × 0.27 × 3.00 0.55 × 0.55 × 2.73 0.55 × 0.55 × 2.73 1.02 × 1.02 × 3.00
Time for acquisition (min:s) 2:48 4:54 3:50 5:16

Note—Multiparametric MRI parameters are given for prostate imaging at 3 T with use of an endorectal coil. DCE = dynamic contrast-enhanced.

a

For apparent diffusion coefficient map calculation. Five evenly spaced b values (0–750 s/mm2) were used.

b

b = 2000 s/mm2.

c

Images obtained before, during, and after administration of single dose of gadopentetate dimeglumine 0.1 mmol/kg at 3 mL/s. Acquisitions performed at 5.6-second intervals.