Abstract
Image quality of PET reconstructions is degraded by subject motion occurring during the acquisition. MR-based motion correction approaches have been studied for PET/MR scanners and have been successful at capturing regular motion patterns, when used in conjunction with surrogate signals (e.g. navigators) to detect motion. However, handling irregular respiratory motion and bulk motion remains challenging. In this work, we propose an MR-based motion correction method relying on subspace-based real-time MR imaging to estimate motion fields used to correct PET reconstructions. We take advantage of the low-rank characteristics of dynamic MR images to reconstruct high-resolution MR images at high frame rates from highly undersampled k-space data. Reconstructed dynamic MR images are used to determine motion phases for PET reconstruction and estimate phase-to-phase nonrigid motion fields able to capture complex motion patterns such as irregular respiratory and bulk motion. MR-derived binning and motion fields are used for PET reconstruction to generate motion-corrected PET images. The proposed method was evaluated on in vivo data with irregular motion patterns. MR reconstructions accurately captured motion, outperforming state-of-the-art dynamic MR reconstruction techniques. Evaluation of PET reconstructions demonstrated the benefits of the proposed method in terms of motion artifacts reduction, improving the contrast-to-noise ratio by up to a factor 3 and achieving a target-to-background ratio up to 90% superior compared to standard/uncorrected methods. The proposed method can improve the image quality of motion-corrected PET reconstructions in clinical applications.
Keywords: PET motion correction, subspace modeling, low-rank reconstruction, PET/MR
1. Introduction
Motion, including physiological motion (i.e., cardiac and respiratory motions) and involuntary bulk motion, is a major source of image quality degradation in Positron Emission Tomography (PET), which can result in spatial blurring artifacts and mismatch between emission and attenuation maps, altering quantification of tracer concentration and deteriorating the diagnostic value of PET images (Liu et al., 2009, Ouyang et al., 2013, Rubeaux et al., 2017). The conventional way to handle motion in PET is the gating method, which bins PET list-mode data to different cardiac and/or respiratory motion phases followed by reconstructions of images of each phase. However, gating results in increased noise levels due to the reduced number of events in each motion phase. To address the limitations of the gating method, many PET motion correction methods have been developed, which consist of two consecutive steps: motion field estimation and motion correction by either applying the estimated motion fields to the gated images or modeling it within motion-compensated PET image reconstruction (Rahmim et al., 2013).
PET motion correction methods can be divided into two major subcategories, depending on how the motion field is estimated: PET-based methods and Magnetic Resonance (MR)-based methods. In the PET-based motion correction methods, the measured emission data are first assigned to specific motion phases based on surrogate signals (Jin et al., 2013), e.g., electrocardiogram (EKG), respiratory bellow, optical tracking, etc. (Fulton et al., 2002, Montgomery et al., 2006, Yu et al., 2016), or the PET-data themselves (Kesner et al., 2009, Sun et al., 2019, Lu et al., 2019), e.g., center of mass, time-of-flight information, frame-by-frame images, etc. Motion fields are then estimated by registering the reconstructed image of each phase to a reference phase (Dawood et al., 2008). However, the accuracy of the motion fields estimated by the PET-based methods is limited by low signal-to-noise ratio (SNR), especially in the case of dual gating, and the overall lack of anatomical structural information of PET images (Ouyang et al., 2013, Petibon et al., 2019).
The increasing availability of hybrid PET/MR systems provides a unique opportunity for mitigating effects of motion in PET using MR-based motion correction. Because of its excellent soft-tissue contrast, high spatial resolution, and high SNR, MR provides more accurate estimation of motion fields than the PET-based methods. MR-based PET motion correction methods have been successfully applied to compensate respiratory and cardiac motion in various applications involving both static and dynamic PET imaging (Petibon et al., 2013, Huang et al., 2014, Petibon et al., 2019, Catana, 2015, Gillman et al., 2017, Küstner et al., 2017). One major limitation of the MR-based motion correction methods is that the conventional noniterative MR imaging methods are unable to resolve cardiac or respiratory motion in real time due to the slow imaging speed. Binning-based MR imaging methods (Grimm et al., 2015, Rank et al., 2016, Feng et al., 2016, Munoz et al., 2018, Robson et al., 2018) are often used to address this issue, where MR k-space data are grouped into different motion phases based on surrogate signals (e.g., EKG), navigator signals, or k-space data alone, and images of each motion phase are then reconstructed for the estimation of motion fields. However, the binning-based MR imaging methods suffer from three noticeable limitations. First, they assume pseudo periodic motion, which does not hold well in the case of arrhythmia and irregular respiratory motion. Second, they rely on either surrogate signals or navigator signals acquired along a single direction to assign k-space data to specific motion phases, which cannot reliably capture involuntary bulk motion. Third, their performance is limited by the inherent trade-off between the number of motion phases (and thus the accuracy of motion field measurement) and data acquisition time.
In this work, we propose a real-time MR imaging method for PET motion corrections in PET/MR. High resolution real-time MR imaging is achieved by a subspace-based imaging method, which takes advantage of a unique property of high-dimensional dynamic MR signals known as partial separability (PS) (Liang, 2007). The PS-model takes advantages of the spatial-temporal correlations of dynamic MR images, significantly reduces the number of unknowns of the underlying spatiotemporal signal, and makes it possible to recover high resolution, high frame-rate dynamic MR images from highly undersampled k-space data (Zhao et al., 2012, Christodoulou et al., 2014). For PET motion correction, the reconstructed real-time MR images are used to determine motion phases and estimate motion fields. PET list-mode data are binned into sinograms accordingly and ordered-subset expectation-maximization (OSEM) reconstruction (Hudson and Larkin, 1994) is performed integrating the estimated displacement in the system matrix for motion correction. We demonstrate the performance of the proposed method by carrying out in vivo 18F-FDG PET/MR imaging experiments using a 3T simultaneous PET/MR scanner.
2. Methods
2.1. PET/MR imaging experiment
An 18F-FDG PET/MR scan was performed on one healthy subject under a study protocol approved by our local IRB. PET and MR data were simultaneously acquired 30 minutes after 18F-FDG injection (around 10 mCi) using a 3T PET/MR scanner (Siemens Biograph mMR, Siemens Healthcare, Erlangen, Germany).
Two 5-minute MR acquisitions were performed using a spoiled gradient-recalled echo (GRE) sequence with stack-of-stars radial sampling trajectories in the coronal plane. The imaging parameters are as follows: image size = 384×384×32, resolution = 1.9 × 1.9 × 5mm3, TR/TE = 3/1.6 ms, and flip angle = 7 degrees. The (k,t)-space data were acquired using a random sampling pattern shown in figure 1. A total of 35 k-space spokes were sampled in each frame, resulting in a frame rate of 9.5 volumes per second. For each frame, the first 3 spokes were respectively acquired along the kx, ky and kz direction across the center of the k-space to estimate the temporal basis of the partially separable (PS) model detailed next. The remaining 32 spokes were along a random angle in the kx−ky plane for every kz. During the first 5-minute acquisition, the subject was instructed to move once to assess the effect of both respiratory and bulk motion. During the second 5-minute acquisition, the subject was instructed to simulate an irregular respiratory pattern including both deep and shallow breaths. The vendor-provided two-point Dixon sequence was performed with breath-holding to obtain attenuation coefficients.
Figure 1:

MR Sampling scheme used for the proposed method. 35 lines per frame are acquired: 1 line along kz and 34 lines in the kx−ky plane. For a better visibility, only the kx−ky in-plane acquired lines are shown in the figure. Two training lines along kx and ky at kz = 0 are consistently acquired through the whole acquisition to estimate the temporal basis Vt (red lines). A random angle is chosen every frame and is consistently acquired every kz for imaging.
2.2. Subspace-based image reconstruction
Denote the dynamic image series as ρ(r, t) and its matrix representation such that:
| (1) |
We express ρ(r, t) as a PS model (Liang, 2007):
| (2) |
or, equivalently C as:
| (3) |
where concentrates in rows the temporal basis function vl of the PS model up to order L and concentrates in columns the corresponding spatial coefficients ul.
We estimate the temporal basis functions using the training data acquired at every frame. Assuming p training lines, we form the so-called Casorati matrix by stacking the signal of the p training lines at each frame, where Nf is the number of samples acquired in each k-space line. The temporal basis functions can then be estimated by calculating the first L right eigenvectors of Ct using Singular Value Decomposition (SVD).
Once Vt is obtained (denoted by ), the image reconstruction problem is reduced to the determination of the spatial coefficients matrix Us. We solve this problem by fitting the PS model to the undersampled (k,t)-space data with additional sparsity constraints (Zhao et al., 2012):
| (4) |
where d is the measured k-space data, Fs is the Fourier transform operator in the spatial domain, i.e. Non-Uniform FFT (NUFFT) operator (Fessler and Sutton, 2003) for the stack-of-stars trajectory, Ω is the sparse sampling operator in the (k,t)-space, T is the finite difference operator along both the spatial and temporal directions, ‖.‖F is the Frobenius norm and the scalar variables λ1 and λ2 are regularization parameters. The first term in Eq. (4) is a data fidelity term, the second term promotes sparsity in the reconstructed image and the third term favors minimal norm solutions for Us.
We solve the optimization problem in Eq. (4) using the Alternating Direction Methods of Multipliers (ADMM) algorithm (Boyd et al., 2011), which leads to solving the following three sub-optimization problem in an alternative fashion:
| (5) |
| (6) |
| (7) |
where z is the split variable, η is the dual variable, and μ is a scalar relaxation parameter. The z update (Eq. (5)) is a soft thresholding operation and the Us update (Eq. (6)) is a convex-optimization problem, which is solved using the conjugate gradient algorithm.
For comparison, we reconstructed MR images using the same data by a binning-based method, known as XD-GRASP (Feng et al., 2016). The respiratory motion signal used for binning was processed the same way Feng et al. did: the Fourier transform of the data at the center of the k-space at each frame was sorted into a 2D matrix, with data from each coil concatenated along the first dimension. A Principal Component Analysis (PCA) was then applied on this matrix and the component with the highest peak in the respiratory frequency range (0.1, 0.5 Hz) was selected as the binning signal. The k-space data were then regrouped into balanced bins, i.e., each bin containing the same number of spokes. The actual number of bins is experiment dependent and will be described later in the Results section. The XD-GRASP reconstruction was performed the same way as in (Feng et al., 2016).
2.3. Motion estimation
The reconstructed real-time MR images were first binned into a small number of phases corresponding to different body positions (respiratory and bulk motion phases). Binning was performed in three steps. The first step consists in visually determining the bulk motion phases from the MR images and discarding time frames corresponding to the transition between bulk motion phases. In the second step, a bin is assigned to each (real-time) frame by tracking the tip of the right lobe of the liver over time while ensuring balanced bins (i.e., all bins should contain a similar number of frames). Finally, a combined MR image is formed for each bin by averaging all real-time images in a bin. Volumetric image registration was then performed between all bins and a reference bin using the multiscale B-spline registration algorithm described in (Chun and Fessler, 2009).
2.4. PET reconstruction
The acquired list-mode PET events were first rearranged into B sinograms y = (y1, …, yB) following the binning determined from MR images and discarding PET list-mode events occurring during bulk motion transitions.
PET reconstruction was performed using the OSEM algorithm (Hudson and Larkin, 1994) integrating the estimated motion fields in the forward model (Liu et al., 2011, Petibon et al., 2016). Let x denote the PET image to reconstruct arranged in vector form. The system matrix, denoted by F is decomposed as F = S AGM, where:
M = [M1, …, MB]⊤ is a stack of deformation operators estimated using the procedure described in Section 2.3,
is a block-diagonal geometrical projection matrix constructed by repeating the static projection matrix implemented using Siddon’s algorithm (Siddon, 1985),
A = diag(A1, …, AB) is a diagonal matrix with time-varying attenuation coefficients,
is a diagonal matrix with detector sensitivity coefficients repeated for all bins.
With these notations, the motion-corrected OSEM update for a given subset l is given by:
| (8) |
where Fl is the system matrix for the l-th subset and sl is the combined additive correction sinogram for subset l including randoms and scatter. Correction sinograms were constructed as follows. Random coincidences were estimated using the delayed window method. Scatter was estimated using the single scatter simulation algorithm (Werling et al., 2002) from an initial reconstruction performed without motion correction. Scatter was estimated separately for each bulk motion phase. Attenuation coefficients were obtained from a vendor-provided Dixon sequence during breath-holding. The attenuation map was deformed to each bin and forward projected to calculate sinogram-domain attenuation coefficients.
The OSEM used 12 subsets and 5 iterations. This motion-corrected reconstruction is denoted by MC in the rest of the paper. For comparison, two other reconstruction methods were considered: a traditional OSEM without motion correction (NMC) and a gated reconstruction where only list-mode events occurring in a given motion phase are reconstructed without motion correction (Gated). Note that no time-of-flight information was available and that the scanner’s point spread function was not included in the imaging model. Both reference methods used 4 iterations to account for the difference in convergence speed, aiming to match the noise level in MC and NMC reconstructions. The noise level was not matched for the Gated method in order to keep gated reconstructions sharp and better compare with motion correction methods (in practice gated reconstruction would typically use fewer subsets and/or iterations).
2.5. Quantitative analysis
In order to compare PET reconstructions, two evaluation measures were used: the contrast-to-noise ratio (CNR) and target-to-background ratio (TBR). The contrast-to-noise ratio is defined as:
| (9) |
where σ0 is the standard deviation in the background region and is the average activity of image x in region . The background region was a spherical region with a diameter of 14 mm, which was placed in a low activity region (lung) and defined large enough to estimate the noise level. With the same notations, the target-to-background ratio is given by:
| (10) |
where is a small spherical region with diameter 4 mm located in the liver. Both measures were evaluated in a small spherical region with diameter 4 mm located in the kidney ( shown in figure 2). The region was selected in the kidney with the highest activity and its size was defined to cover the high activity region. Note that all regions of interest were located at the same position for all reconstruction methods. In the absence of ground truth, the contrast-to-noise and target-to-background ratios were used as indicators of image quality.
Figure 2:

Regions of interest used for quantitative analysis: (a) bulk motion experiment, (b) irregular respiratory motion experiment. ROI 0 in the lung, ROI 1 in the kidney, ROI 2 in the liver. For each experiment a small region in the kidney with the highest activity was selected to assess the effects of different PET reconstruction methods on contrast-to-noise ratio.
3. Results
3.1. Correction of bulk motion
In this experiment the subject was instructed to move after around 2.5 minutes in the 5-minute acquisition. Images reconstructed by XD-GRASP and the proposed method at the end-inhalation and end-exhalation phases are shown in figure 3. The k-space data were grouped into 6 balanced bins as in (Feng et al., 2016) in the XD-GRASP reconstruction. The images obtained by XD-GRASP method show noticeable blurring artifacts largely because the bulk motion was not detected from the navigator signal. More specifically, figure 4(a) shows the navigator signal obtained from the training line along the kz direction in each frame as in (Feng et al., 2016). Since the bulk motion of the subject was along the x direction (left to right), the navigator signal only recorded abnormal changes during the bulk motion, i.e., the red region in figure 4(a), but did not contain sufficient information to indicate what type of motion occurred. Therefore, six motion bins were chosen in XD-GRASP while the k-space data acquired in the red region of figure 4(a) were discarded, resulting in blurring artifacts.
Figure 3:

Representative MR images obtained with the XD-GRASP and the proposed method. The red dashed-line indicates the top of the liver position for end-inhalation and the green dashed-line indicates the top of the liver position for end-exhalation. Note that, besides respiratory motion, bulk motion indicated by the yellow arrows is clearly seen in the proposed low-rank based image reconstruction. The images obtained by XD-GRASP show blurring artifacts largely because the bulk motion was not detected from the navigator signal (see figure 4 for more details).
Figure 4:

(a) Plot of the processed navigator along time, which was used to bin the k-space data in XD-GRASP. (b) to (d) Real part of the temporal basis Vt for the component 1, 2, and 15 of the PS model, respectively.
The images obtained by the proposed method shown in figure 3 successfully capture both respiratory motion (as indicated by the red and green dashed lines) and bulk motion (as indicated by the yellow arrows). Figure 4(b) to (d) show the temporal basis functions of the PS model estimated from the three training lines, where, intuitively, the first component (figure 4(b)) shows respiratory patterns and the second component (figure 4(c)) indicates bulk motion. To further demonstrate the real-time capabilities of the proposed method, figure 5 shows images at multiple time frames along with a 1D profile through the liver along time. The images from before and after bulk motion demonstrate the ability to capture both respiratory and bulk motion. The yellow overlay emphasizes the body displacement between Stage 1 and Stage 2. The profile plot shows the respiratory motion, captured for both bulk motion phases. The transition portion between the two bulk motion phases corresponds to the frames that were discarded in the PET reconstruction. Videos showing the reconstructed MR images are available in the Supplementary Material M1.
Figure 5:

Dynamic MR images reconstructed by the proposed method. The real-time profile (e) is plotted for a part of the experiment where the bulk motion happens, and two representative images (a) & (b) and (c) & (d) are shown for each body position. The white line in (a) shows where the time profile was taken. A yellow box at the edge of the patient has been drawn for the first body position (b) and the same box was also drawn for the second body position (d) at the same coordinates (regarding the image). One can clearly see that the body of the subject moved to the right of the image during the bulk motion, and that the proposed method managed to catch that motion.
To account for the two body positions and for respiratory motion in PET reconstruction, real-time MR images obtained by the proposed method were grouped into 12 bins (6 bins for each body position) for motion field estimation. Frames in the transition between the two bulk motion phases were excluded (a total of 15 seconds were discarded). Motion was estimated between all bins and the bin corresponding to the end-exhalation, which was used as reference bin. Estimated motion fields are shown in figure 6. The top left image shows motion caused by respiration, mostly visible as a vertical displacement near the liver (shown with more details in the inset image). The left column images show the bulk motion, which is mostly lateral. Finally, the bottom right figure shows a combination of respiratory motion between end-inhalation and end-exhalation and bulk motion.
Figure 6:

Estimated motion field between bins from different bulk motion phases. The top left panel corresponds to the bin used as reference (Body position 1 and the end of exhalation). The top right panel shows the same body position at the end of inhalation; the overlaid motion field exhibits mostly vertical displacement near the liver, corresponding to respiratory motion. The bottom row shows the estimated motion field at the end of exhalation and inhalation for the second body position (after bulk motion). Motion fields demonstrate the lateral displacement between body positions.
Reconstructed PET images are shown in figure 7. Figures 7(a) and 7(b) show coronal and axial slices using different reconstruction methods. Motion caused blurring artifacts are clearly visible on the NMC reconstruction primarily in the lateral direction, corresponding to bulk motion but also in the vertical direction due to respiratory motion. The gated reconstruction, which uses one sixth of the PET counts at a single body position, shows sharper features but is severely corrupted by noise. The motion-corrected reconstruction using XD-GRASP for motion estimation (denoted by MC-XDG) still shows noticeable motion artifacts because the XD-GRASP reconstruction failed to capture the bulk motion. The proposed method compensates both respiratory and bulk motion, significantly reducing motion-blur, while exhibiting a low noise level. Figure 7(c) shows line profiles through the kidney. Without motion correction, the activity peak is lowered by motion. Instead, two distinct peaks are visible, which correspond to the two bulk motion phases. The two peaks are also visible for the XD-GRASP based motion correction. The gated reconstructions and the motion-corrected reconstructions by the proposed method both preserve the peak activity. However, the gated reconstructions exhibit a high level of noise due to the reduced amount of data used for reconstruction.
Figure 7:

PET reconstructions for the bulk motion experiment using three different methods: reconstruction without motion correction (NMC), reconstruction from PET data corresponding to a single respiratory phase and body position (Gated), motion correction using motion estimated from the XD-GRASP MR reconstructions (MC-XDG) and proposed motion-corrected reconstruction from low-rank MR reconstruction (MC-LR). Profile plots through the right kidney (along the orange line drawn on the NMC image) are shown in (c).
Evaluation measures are reported in table 1. The table shows that motion corrected reconstruction leads to the highest CNR: a 83% improvement was observed over reconstruction without motion correction, 47% over XD-GRASP-based motion correction and 198% over gated reconstruction. The noise level in NMC and MC is similar (within 15%) but the contrast is substantially improved by motion correction, while the noise level in the gated reconstruction is three times higher leading to the low CNR. This is expected since gated reconstructions were not tuned to reduce noise but to increase contrast. One could expect a lower CNR for gated reconstruction with a tuned number of subsets and iterations. For the TBR, the gated reconstruction achieves the highest ratio, because gated reconstructions favor high contrast (at the expense of high noise). The proposed motion compensation method approaches the gated TBR (25% decrease) and outperforms NMC (20% increase).
Table 1:
Contrast-to-noise ratio (CNR) and target-to-background ratio (TBR) for kidney region of interest. See figure 2 for a view of the regions of interest.
| NMC | Gated | MC-XDG | MC-LR | |
|---|---|---|---|---|
| CNR | 42.07 | 25.88 | 52.49 | 77.26 |
| TBR | 6.05 | 9.57 | 6.84 | 7.25 |
3.2. Correction of irregular respiratory motion
The second experiment was designed to evaluate the performance of the proposed method in the case of irregular respiratory motion. The subject was instructed to alternate between slow deep and fast shallow breaths throughout the 5 minutes PET/MR acquisition.
MR images obtained by the proposed method are shown in figure 8. The top row shows images at different frames: two at the end of inhalation and two at the end of exhalation taken from different breathing patterns (deep/shallow), respectively. The full extent of the respiratory motion is captured and the images are artifacts-free. Figures 8(e) and 8(f) show 1D profiles of the image through the liver changing over time. Both the images and the plot in figure 8 clearly show the breathing patterns, alternating between deep slow breaths and fast shallow ones. Based on the reconstructed real-time MR images, 12 bins were determined through analysis of the liver displacement in the MR images and were consequently used for motion field estimation and motion corrected PET reconstruction. Sequences of MR reconstructions are shown in Supplementary Material M2. Reconstructed MR images using XD-GRASP are shown in Supplementary Material M3, where the k-space data were grouped into 12 balanced bins to capture both deep and shallow respiratory motion.
Figure 8:

Reconstructed MR images in the case of irregular respiratory motion. The real-time profile (e) is plotted for a few minutes of the experiment where the subject changes their breathing pattern from slow and deep to fast and shallow respiration. Images (a) and (b) show representative images at end-inhalation during deep breath and shallow breath, respectively; (c) and (d) at end-exhalation. The red dashed-line indicates the top of the liver position for end-inhalation and the green dashed-line indicates the top of the liver position for end-exhalation for each breathing pattern.
Corresponding PET reconstructions are shown in figure 9. Images reconstructed without motion correction (NMC) exhibit blurring artifacts. This is particularly visible on the left kidney (see the green line on the gated coronal image) where the bright spot visible on other images is elongated in the vertical direction, due to the large amplitude of the respiratory motion. The gated reconstruction uses one sixth of the total number of counts and therefore is degraded by noise, despite resulting in a sharper image. The motion corrected reconstructions using XD-GRASP and the proposed method for motion estimation both produce sharper images with higher SNR, while the proposed motion correction method results in the best image quality, in terms of noise and contrast. Corresponding line profiles are plotted in figure 9(c). The NMC peak is elongated along the y-axis, due to the large extent of the mostly vertical respiratory motion. The Gated line profile is sharper near its peak but has a large noise level. The proposed MC method results in a good compromise between sharpness and low noise. Contrast-to-noise and target-to-background ratios (defined in Eq. (9) and Eq. (10) respectively) are reported in table 2 (regions of interest are shown in figure 2(b)). Metrics show the superior performance of the proposed motion correction method. The improvement in CNR is around 163% over NMC and over 200% over gated reconstruction. As for the bulk motion experiment the gated reconstructions were tuned to maximize contrast rather than reduce noise. The CNR for the gated reconstructions could be improved with further tuning (i.e. by decreasing the number of subsets and iterations). The TBR for the proposed method is within 15% of the gated TBR and around 95% larger than NMC.
Figure 9:

PET reconstructions for the irregular motion experiment using three different methods: reconstruction without motion correction (NMC), reconstruction from PET data corresponding to a single respiratory phase (Gated) and proposed motion-corrected reconstruction (MC-LR). (c) shows profile plots through the left kidney (the profile line is shown on the NMC coronal PET image).
Table 2:
Contrast-to-noise ratio and target-to-background ratio for kidney region of interest. Regions are shown in figure 2.
| NMC | Gated | MC-XDG | MC-LR | |
|---|---|---|---|---|
| CNR | 17.52 | 15.28 | 37.39 | 46.18 |
| TBR | 3.92 | 9.00 | 6.33 | 7.64 |
4. Discussions
We have demonstrated the performance of the proposed MR-based motion correction for PET in two challenging cases: bulk motion and irregular respiratory motion. The proposed subspace-based MR imaging method allows for reconstruction of high-resolution 3D volumes at a rate of 9.5 volume/s, which enables accurate motion field estimation even in the case of irregular motions. Another important benefit of the proposed approach is the ability to perform informed binning for PET motion correction, rather than relying on navigators or external markers which offer limited information on the subject motion. With full real-time volumetric MR images, detecting motion becomes straightforward, and the process of determining an appropriate number of bins is greatly simplified.
The key assumption of the subspace-based imaging method is the low-rank property of dynamic MR signals. We performed a simulation study to investigate this property in the case of regular and irregular respiratory motion. Two phantoms (shown in the Supplemental Material M4 and M5) were generated using the XCAT software (Segars et al., 2010) to simulate regular and irregular respiratory motion. Respiratory and cardiac cycles were divided into respectively 30 and 40 phases and 3D volumes were computed for each respiratory and cardiac phase combination (i.e. 1200 volumes). A 4D (3D space + time) phantom was then built by selecting and concatenating frame by frame the 3D volumes based on simulated EKG and respiratory signals. Both the breathing frequency and diaphragm expansion were varied while keeping a constant heart rate in the simulation of the irregular respiratory motion. Each phantom contained 6 respiratory cycles. The contrast was designed to simulate a Balanced Steady-State Free Precession (bSSFP) signal for several compartments such as fat, muscles, etc. using T1 and T2 values from the literature (Bojorquez et al., 2017). SVD was then performed to investigate the effect of an irregular respiratory pattern on the rank, and its corresponding approximation error with low-rank truncation (see Supplemental Material figure M6). The decay of the calculated singular values from both phantoms was very similar, indicating that the breathing pattern does not substantially affect the rank of the data.
The proposed method utilizes an MR acquisition which fully overlaps with the PET acquisition and provides real-time MR images for motion correction. The proposed method can still have benefits for other commonly used acquisition protocols. It is common in practice to reserve a first part of the PET acquisition to perform MR motion field measurements and use the remaining PET acquisition time to perform additional MR measurements (e.g. using T1 or T2 contrast sequences) that can be used for other diagnostic tasks (Petibon et al., 2019). The proposed method can advantageously replace the motion field measurement sequence, possibly reducing the acquisition time while preserving image quality. A gating signal (e.g. navigator or external marker) can then be used in subsequent MR sequences to select an appropriate bin for each PET frame. Another approach is to integrate contrast sequences into the motion field estimation sequence described in this paper. This is under investigation and will be reported in separate publications.
The study reported in this paper has several limitations. First, the computation time for the low-rank reconstruction with sparsely sampled non-Cartesian k-space data could be a concern. The current MATLAB (The MathWorks, Inc., Natick, Massachusetts, United States) implementation performs reconstruction of one slice and one coil in around one hour. We anticipate that using a lower level programming language and parallel computing devices (e.g. GPU) will help achieve reasonable runtimes (Wu et al., 2011). Second, the proposed method does not have sufficient temporal resolution to resolve the motion in the transition phase between the two bulk motion phases of experiment 1 (figure 5(e)). The time-varying profile plot shows that the image quality in the transition is severely degraded. The corresponding list-mode data were excluded from the PET reconstruction. Since the duration of the bulk motion was short, only about 5% of the list-mode data were discarded and thus should not be a significant limitation. Third, this study focuses on demonstrating the feasibility of using subspaced-based real-time MR for PET motion correction. We showed the performance of our method in two cases (bulk motion and irregular respiratory motion) from in vivo PET/MR experiments on a healthy subject. More subjects are needed to fully evaluate the performance of the proposed method in clinical settings. It is also worthwhile to explore incorporating advanced system modeling including time-of-flight information and point-spread function modeling into the proposed method. While advanced imaging models are expected to improve low-count (e.g. gated) reconstructions and quantitative metrics, all reconstruction methods, including the proposed motion correction, would benefit from more accurate modeling.
5. Conclusion
We proposed an MR-based method for PET motion correction using a subspace-based real-time MR imaging for motion field estimation. We demonstrate the feasibility of the proposed method using 18F-FDG-PET/MR studies on a healthy subject. Our results show that the proposed method can capture and correct for normal and irregular respiratory motions as well as bulk body motion. The proposed method can be beneficial to a range of clinical applications where irregular motion patterns are expected.
Supplementary Material
Acknowledgments
This work was supported in part by the National Institutes of Health under award numbers: T32EB013180, R01CA165221, R01HL118261, R21MH121812, R01HL137230 and P41EB022544.
References
- Bojorquez JZ, Bricq S, Acquitter C, Brunotte F, Walker PM and Lalande A 2017. What are normal relaxation times of tissues at 3T? Magnetic Resonance Imaging 35, 69–80 [DOI] [PubMed] [Google Scholar]
- Boyd S, Parikh N, Chu E, Peleato B and Eckstein J 2011. Distributed Optimization and Statistical Learning via the Alternating Direction Method of Multipliers Foundations and Trends in Machine Learning 3(1), 1–122 [Google Scholar]
- Catana C 2015. Motion correction options in PET/MRI Seminars in Nuclear Medicine 45(3), 212–223 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Christodoulou AG, Hitchens TK, Wu YL, Ho C and Liang ZP 2014. Improved subspace estimation for low-rank model-based accelerated cardiac imaging IEEE Transactions on bio-Medical Engineering 61(9), 2451–2457 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chun SY and Fessler JA 2009. A simple regularizer for B-spline nonrigid image registration that encourages local invertibility IEEE Journal of Selected Topics in Signal Processing 3(1), 159–169 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dawood M, Buther F, Jiang X and Schafers KP 2008. Respiratory motion correction in 3-D PET data with advanced optical flow algorithms IEEE Transactions on Medical Imaging 27(8), 1164–1175 [DOI] [PubMed] [Google Scholar]
- Feng L, Axel L, Chandarana H, Block KT, Sodickson DK and Otazo R 2016. XD-GRASP: Golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing Magnetic Resonance in Medicine 75(2), 775–788 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fessler JA and Sutton BP 2003. Nonuniform fast Fourier transforms using min-max interpolation IEEE Transactions on Signal Processing 51(2), 560–574 [Google Scholar]
- Fulton RR, Meikle SR, Eberl S, Pfeiffer J, Constable CJ and Fulham MJ 2002. Correction for head movements in positron emission tomography using an optical motion-tracking system IEEE Transactions on Nuclear Science 49(1), 116–123 [Google Scholar]
- Gillman A, Smith J, Thomas P, Rose S and Dowson N 2017. PET motion correction in context of integrated PET/MR: Current techniques, limitations, and future projections Medical Physics 44(12), e430–e445 [DOI] [PubMed] [Google Scholar]
- Grimm R, Furst S, Souvatzoglou M, Forman C, Hutter J, Dregely I, Ziegler SI, Kiefer B, Hornegger J, Block KT and Nekolla SG 2015. Self-gated MRI motion modeling for respiratory motion compensation in integrated PET/MRI Medical Image Analysis 19(1), 110–120 [DOI] [PubMed] [Google Scholar]
- Huang C, Ackerman JL, Petibon Y, Normandin MD, Brady TJ, El Fakhri G and Ouyang J 2014. Motion compensation for brain PET imaging using wireless MR active markers in simultaneous PET-MR: phantom and non-human primate studies NeuroImage 91, 129–137 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hudson HM and Larkin RS 1994. Accelerated image reconstruction using ordered subsets of projection data IEEE Transactions on Medical Imaging 13(4), 601–609 [DOI] [PubMed] [Google Scholar]
- Jin X, Mulnix T, Gallezot JD and Carson RE 2013. Evaluation of motion correction methods in human brain PET imaging–a simulation study based on human motion data Medical Physics 40(10), 102503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kesner AL, Bundschuh RA, Detorie NC, Dahlbom M, Ziegler SI, Czernin J and Silverman DH 2009. Respiratory Gated PET Derived in a Fully Automated Manner From Raw PET Data IEEE Transactions on Nuclear Science 56(3), 677–686 [Google Scholar]
- Küstner T, Schwartz M, Martirosian P, Gatidis S, Seith F, Gilliam C, Blu T, Fayad H, Visvikis D, Schick F, Yang B, Schmidt H and Schwenzer NF 2017. MR-based respiratory and cardiac motion correction for PET imaging Medical Image Analysis 42, 129–144 [DOI] [PubMed] [Google Scholar]
- Liang ZP 2007. Spatiotemporal Imaging with Partially Separable Functions in ‘IEEE International Symposium on Biomedical Imaging’ 2007 4th IEEE International Symposium on Biomedical Imaging: From Nano to Macro pp. 988–991 [Google Scholar]
- Liu C, Alessio AM and Kinahan PE 2011. Respiratory motion correction for quantitative PET/CT using all detected events with internal-external motion correlation Medical Physics 38(5), 2715–2723 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu C, Pierce L A n, Alessio AM and Kinahan PE 2009. The impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging Physics in Medicine and Biology 54(24), 7345–7362 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lu Y, Gallezot JD, Naganawa M, Ren S, Fontaine K, Wu J, Onofrey JA, Toyonaga T, Boutagy N, Mulnix T, Panin VY, Casey ME, Carson RE and Liu C 2019. Data-driven voluntary body motion detection and non-rigid event-by-event correction for static and dynamic PET Physics in Medicine and Biology 64(6), 065002. [DOI] [PubMed] [Google Scholar]
- Montgomery AJ, Thielemans K, Mehta MA, Turkheimer F, Mustafovic S and Grasby PM 2006. Correction of head movement on PET studies: comparison of methods Journal of Nuclear Medicine 47(12), 1936–1944 [PubMed] [Google Scholar]
- Munoz C, Neji R, Cruz G, Mallia A, Jeljeli S, Reader AJ, Botnar RM and Prieto C 2018. Motion-corrected simultaneous cardiac positron emission tomography and coronary MR angiography with high acquisition efficiency Magnetic Resonance in Medicine 79(1), 339–350 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ouyang J, Li Q and El Fakhri G 2013. Magnetic resonance-based motion correction for positron emission tomography imaging Seminars in Nuclear Medicine 43(1), 60–67 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Petibon Y, Guehl NJ, Reese TG, Ebrahimi B, Normandin MD, Shoup TM, Alpert NM, El Fakhri G and Ouyang J 2016. Impact of motion and partial volume effects correction on PET myocardial perfusion imaging using simultaneous PET-MR Physics in Medicine and Biology 62(2), 326–343 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Petibon Y, Ouyang J, Zhu X, Huang C, Reese TG, Chun SY, Li Q and El Fakhri G 2013. Cardiac motion compensation and resolution modeling in simultaneous PET-MR: a cardiac lesion detection study Physics in Medicine and Biology 58(7), 2085–2102 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Petibon Y, Sun T, Han PK, Ma C, El Fakhri G and Ouyang J 2019. MR-based cardiac and respiratory motion correction of PET: application to static and dynamic cardiac (18)F-FDG imaging Physics in Medicine and Biology 64(19), 195009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rahmim A, Tang J and Zaidi H 2013. Four-Dimensional Image Reconstruction Strategies in Cardiac-Gated and Respiratory-Gated PET Imaging PET Clinics 8(1), 51–67 [DOI] [PubMed] [Google Scholar]
- Rank CM, Heußer T, Wetscherek A, Freitag MT, Sedlaczek O, Schlemmer HP and Kachelrieß M 2016. Respiratory motion compensation for simultaneous PET/MR based on highly undersampled MR data Medical Physics 43(12), 6234. [DOI] [PubMed] [Google Scholar]
- Robson PM, Trivieri MG, Karakatsanis NA, Padilla M, Abgral R, Dweck MR, Kovacic JC and Fayad ZA 2018. Correction of respiratory and cardiac motion in cardiac PET/MR using MR-based motion modeling Physics in Medicine and Biology 63(22), 225011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rubeaux M, Doris MK, Alessio AM and Slomka PJ 2017. Enhancing Cardiac PET by Motion Correction Techniques Current Cardiology Reports 19(2), 14. [DOI] [PubMed] [Google Scholar]
- Segars WP, Sturgeon G, Mendonca SB, Grimes JK and Tsui BMW 2010. 4D XCAT phantom for multimodality imaging research Medical Physics 37(9), 4902–4915 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Siddon RL 1985. Fast calculation of the exact radiological path for a three-dimensional CT array Medical Physics 12(2), 252–255 [DOI] [PubMed] [Google Scholar]
- Sun T, Petibon Y, Han PK, Ma C, Kim SJW, Alpert NM, El Fakhri G and Ouyang J 2019. Body motion detection and correction in cardiac PET: Phantom and human studies Medical Physics 46(11), 4898–4906 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Werling A, Bublitz O, Doll J, Adam LE and Brix G 2002. Fast implementation of the single scatter simulation algorithm and its use in iterative image reconstruction of PET data Physics in Medicine and Biology 47(16), 2947–2960 [DOI] [PubMed] [Google Scholar]
- Wu XL, Gai J, Lam F, Fu M, Haldar JP, Zhuo Y, Liang ZP, Hwu WM and Sutton BP 2011. Impatient MRI: Illinois Massively Parallel Acceleration Toolkit for image reconstruction with enhanced throughput in MRI in ‘IEEE International Symposium on Biomedical Imaging’ IEEE pp. 69–72 [Google Scholar]
- Yu Y, Chan C, Ma T, Liu Y, Gallezot JD, Naganawa M, Kelada OJ, Germino M, Sinusas AJ, Carson RE and Liu C 2016. Event-by-Event Continuous Respiratory Motion Correction for Dynamic PET Imaging Journal of Nuclear Medicine 57(7), 1084–1090 [DOI] [PubMed] [Google Scholar]
- Zhao B, Haldar JP, Christodoulou AG and Liang ZP 2012. Image reconstruction from highly undersampled (k,t)-space data with joint partial separability and sparsity constraints IEEE Transactions on Medical Imaging 31(9), 1809–1820 [DOI] [PMC free article] [PubMed] [Google Scholar]