Table 1.

Concentration Response of the Calibration Curve and Accuracy of Measuring GMAb Concentration in Human Serum Using the GMAb ELISA.

Expected concentration, ng/ml Measured concentration, ng/ml (Mean ± SD) Coefficient of variation (100×SD/Mean) Standard error (%) (Mean ± SD)
Calibration curve*
 50 51.9 ± 2.3 4.39 4.3 ± 3.9
 25 24.4 ± 0.7 2.79 2.5 ± 2.7
 12.5 12.2 ± 0.4 5.07 4.0 ± 3.5
 6.25 5.8 ± 0.4 6.09 7.7 ± 5.2
 3.125 3.1 ± 0.2 6.35 5.3 ± 2.5
 1.5625 1.3 ± 0.1 10.58 14.8 ± 7.0
 0.78125 0.8 ± 0.2 19.41 16.7 ± 12.1
Spiked samples
 10 10.1 ± 0.6 6.32 5.4 ± 2.9
 20 19.0 ± 1.1 5.89 6.2 ± 4.0
 40 41.8 ± 1.7 4.18 4.5 ± 4.2
*

Prepared using the autoimmune PAP-patient derived PCRS by serial dilution in sample dilution buffer.

Spiked samples were dilution buffers spiked with known concentration of purified GMAb.