Table 1.

Characteristics and main outcome measures of the tested machines

Dialysis machine characteristics F4008H F5008S BB
Number of machines 18 18 31
Number of dialysis treatments 31 53 69
Conductivity (mS/cm) 14.29 ± 0.42 14.05 ± 0.35 14.25 ± 0.42
Programmed DNa (mEq/L) 140.3 ± 3.7 140.3 ± 3.8 139.9 ± 3.8
Measured DNa (mEq/L)
 ISE 145.75 ± 4.8 143.5 ± 3.05 146.5 ± 4.6
 FP 143.9 ± 4.6 140.7 ± 2.9 146.7 ± 4.3
 IC 143.6 ± 4 141.4 ± 3.5 145.4 ± 3.9
Measured minus programmed DNa (mEq/L), median (IQR)
 ISE 5.5 (5−7) 4 (2−5) 7 (6−8)
 FP 4 (2.25−5.75) 1 (−1−1.75) 7 (6−8)
 IC 4 (2−5) 1 (−0.5−2) 6 (5−7)
Measured potassium (mEq/L), median (IQR) NA 2.2 (2.1–2.3) 1.4 (1.1–3.6)
Measured calcium (mmol/L), median (IQR) NA 2.7 (1.7–3.7) 0.92 (0.36-1.52)
Measured magnesium (mmol/L), median (IQR) NA 1.12 (1.06–1.26) 1.05 (0.98–1.14)

Results expressed as mean ± SD unless stated otherwise. 1.0 mEq/L sodium/potassium = 1.0 mmol/L, 2.0 mEq/L calcium/magnesium = 1.0 mmol/L.

NA, not available.