Table 2.

The difference between measured and programmed DNa at 135 and 136 mEq/L

Na measurement method At programmed DNa 135 mEq/L*
At programmed DNa 136 mEq/L**
F4008H F5008S (n = 10) BB (n = 10) F4008H (n = 10) F5008S (n = 4) BB (n = 11)
ISE measured minus programmed DNa NA 6 (4.75−7.25) 7.5 (7−9) 5.5 (4.75−8.75) NA 6 (6−7)
FP measured minus programmed DNa NA 4 (1.75−5) 7 (6−7) 4.5 (0.75−6) NA 8 (7−8)
IC measured minus programmed DNa NA 4.5 (1.75−5.25) 6 (6−7) 5 (3.75−5) 2.5 (2−4.5) 5 (4.75−5.25)

Results expressed as median (IQR).

*

P < 0.05 between F5008S/BB when DNa measured by all methods.

**

P = 0.004 between F4008H/BB when DNa measured by FP.

NA, not available.