Table 3.

The difference between measured and programmed DNa at a programmed DNa of 140 or 145 mEq/L

Na measurement method At prescribed DNa 140 mEq/L*
At prescribed DNa 145 mEq/L**
F4008H (n = 11) F5008S (n = 21) BB (n = 20) F4008H (n = 10) F5008S (n = 17) BB (n = 20)
ISE measured minus programmed DNa 5 (3.5−6) 3 (1−4) 7 (6−8) 7 (1−8.5) 2.5 (0.75−4) 8 (6.5−9)
FP measured minus programmed DNa 4 (1−5.5) 0 (−1−1) 6 (5−7) 4 (3.5−5.5) −0.5 (−2.25−1) 7 (6−8.5)
IC measured minus programmed DNa 2 (2−4) 1 (−1−2) 6 (3.25−6) 4.5 (0.25−5.25) 1 (−1.5−1.5) 6 (4−7)

Results expressed as median (IQR).

*

P < 0.001 between F5008S/BB by ISE, FP and IC.

**

P < 0.001 between F5008S/BB by ISE, FP and IC; P = 0.045 between F4008H/F5008S by IC.