Table 4.

Energy resolution

KETEK SensL Hamamatsu
PA3325-WB-0808 Array-J-30020-64P-PCB S14161-3050-HS-08
Threshold vth_t1
10 10.35±0.02% 11.20±0.02% 13.10±0.02%
20 10.44±0.02% 10.92±0.02% 13.15±0.03%
30 10.84±0.02% 11.22±0.02% 13.12±0.03%
40 10.85±0.02% 11.06±0.02% 13.08±0.03%
50 11.08±0.02% 11.23±0.02% 13.13±0.03%
Segmentation layer
BaSO4 (360 µm) 11.08±0.02% 11.23±0.02% 13.13±0.03%
BaSO4 (110 µm) 12.60±0.02% 11.64±0.02% 14.70±0.02%
ESR (67 µm, glued) 12.58±0.02% 13.78±0.02% 14.74±0.02%
Distance
18 mm 10.94±0.01% - -
38 mm 11.08±0.02% 11.23±0.02% 13.13±0.03%
58 mm 10.67±0.02% 10.66±0.02% 11.39±0.03%
Clinical (19 mm height)
ESR (155 µm, air-coupled) 10.74±0.05% 10.72±0.04% 10.46±0.08%

Parameter study for multi-channel configurations featuring SiPMs by different vendors. Table reports the energy resolution corresponding to the lowest CRTs achieved in coincidence measurements of two times 64 channels (see Table 2). If not indicated differently, data were acquired at vth_t1=50, vth_t2=20, vth_e=15, with 360 µm BaSO4 for scintillator segmentation, a scintillator height of 12 mm, and at a distance of 38 mm. The overvoltage setting varies according to the parameter setting and SiPM type used. For the clinical configuration (19-mm high scintillator), only two times 16 channels were read out