Skip to main content
. Author manuscript; available in PMC: 2022 Mar 2.
Published in final edited form as: IEEE Trans Med Imaging. 2021 Mar 2;40(3):1021–1031. doi: 10.1109/TMI.2020.3044530

Fig. 6.

Fig. 6.

2D search vs. 3D search results. Proportion correct for human observers, ideal observer (IO), non-prewhitening with eye filter (NPWE), channelized Hotelling observer (CHO) using Gabor, Difference of Gaussian, and Laguerre-Gauss channels and foveated search model (FSM) with human eye movements and MAP eye movement model for microcalcification (left) and mass (right). Model observers inside the gray area had their signal contrast reduced to match 3D human performance. Error bars are standard error of the mean across observers, with a confidence interval of 1 standard deviation. Significance is only indicated for human observers’ results. ** = p < 0.01.