Skip to main content
. 2021 Mar 10;8(3):46. doi: 10.3390/vetsci8030046

Table 1.

Validation parameters for the determination of Aflatoxin M1 (AFM1) and Aflatoxin M2 (AFM2) in milk prior to the High-Performance Liquid Chromatography with Fluorescence Detector (HPLC-FL) method.

Compound Linearity
R2
LOD 1
(ng kg−1)
Intra-Day
Recovery (%)
RSD 2 (%)
Inter-Day
Recovery (%)
RSD (%)
CCα 3
(ng kg−1)
CCβ 4
(ng kg−1)
Error α Error β MRL 5
(ng kg−1)
AFM1 0.999 11.99 90–100
< 3.6
96–109
< 17
56.52 63.97 6.52 6.97 50
AFM2 0.996 16.95 91–119
< 6.1
74–120
< 10.6
57.27 65.57 7.27 8.57

1 Limit Of Detection; 2 Relative Standard Deviation; 3 Decision Limit; 4 Detection Capability; 5 Maximum Residue Limits.