MRE protocol. Synoptic table summarizes the scanning parameters of MR pulse sequences. Coronal T1-weighted 3D gradient-echo liver acquisition with volume acceleration (LAVA) sequences with fat suppression were acquired before and after intravenous administration of paramagnetic contrast material (0.1 mmol/kg, followed by 20 mL of saline solution, both at a flow rate of 2 mL/s); in particular, the sequence was acquired at 60 and 120 s after contrast administration. An anti-peristaltic agent, scopolamine-N-butyl bromide (Buscopan® 20 mg; Boehringer Ingelheim, Milano, Italy), was administered intravenously just before post-contrast imaging to reduce peristaltic artifacts.

MRI Protocol Coronal T2W SSFSE Coronal FIESTA Coronal T2W Thick-slab SSFSE Axial DWI SE EPI Sagittal, Coronal, Axial T1W 3D GRE LAVA
Repetition time/Echo time (ms) 705/90 4/1.7 2408/1103 3000/74 4.1/1.9
Flip angle 90° 75° 90° 90° 12°
Section Thickness (mm) 6 6 70 8 3.4
Interslice gap (mm) 0.6 0.6 2 1.7
Bandwidth (kHz) 83.33 100 31.25 250 62.5
Field of view (cm) 44–48 44–48 44–48 42 44–48
Matrix 384 × 224 512 × 384 512 × 384 160 × 160 320 × 192
N. of signal acquired 0.57 1 1 2 0.7
N. of images 28 28 1 15 120
Frequency direction Right to left Right to left Right to left Anterior to posterior Superior to inferior
Acquisition time (s) 24 22 2 27 23
B-value (s/mm2) 0–800

T2W = T2-weighted, T1W = T1-weighted, SSFSE = single-shot fast spin-echo, FIESTA = fast imaging employing steady-state acquisition, DWI = diffusion-weighted imaging, SE = spin-echo, EPI = echoplanar imaging, GRE = gradient-echo, LAVA = liver acquisition with volume acceleration.