Table 2.
Evaluation Results of Test Equipment (GenBody COVID-19 Kit) IgM/IgG | Confirmed Results through RT-PCR | Total | |
---|---|---|---|
Positive | Negative | ||
Positive | 127 | 0 | 127 |
Negative | 3 | 100 | 103 |
Total | 130 | 100 | 230 |
Clinical sensitivity: (127/130) × 100 = 97.69% (95% confidence interval (CI): 93.40–99.52%). Clinical specificity: (100/100) × 100 = 100% (95% CI: 96.38–100%). RT-PCR, SARS-CoV-2 reverse-transcription polymerase chain reaction.