Table 3.
Evaluation Results of Test Equipment (GenBody COVID-19 kit) IgM | Confirmed Results through RT-PCR | Total | |
---|---|---|---|
Positive | Negative | ||
Positive | 107 | 0 | 107 |
Negative | 23 | 100 | 123 |
Total | 130 | 100 | 230 |
Clinical sensitivity: (107/130) × 100 = 82.31% (95% CI: 74.65–88.44%). Clinical specificity: (100/100) × 100 = 100% (95% CI: 96.38–100%). RT-PCR, SARS-CoV-2 reverse-transcription polymerase chain reaction.