Skip to main content
. 2020 Jun 10;105(4):526–530. doi: 10.1136/bjophthalmol-2020-316817

Table 4.

Reliability analysis

Device Sensibility
(95% CI)
Specificity
(95% CI)
Positive predictive value
(95% CI)
Negative predictive value
(95% CI)
Spectralis OCTA 43.3%
(39.1–47.5%)
54.5%
(45.2–63.6%)
81.2%
(77.7–84.3%)
17.4%
(15.1–20.1%)
PlexElite 27.8%
(24.1–31.8%)
69.4%
(60.4–77.5%)
80.5%
(75.4–84.8%)
17.4%
(15.7–19.4%)
RTVue XR 24.2%
(20.7–28%)
69.4%
(60.4–77.5%)
78.2%
(72.6–83%)
16.7%
(15.1–18.6%)
AngioPlex 20.9%
(17.6–24.5%)
71.9%
(63–79.7%)
77.1%
(70.9–82.4%)
16.1%
(15.1–18.4%)
DRI OCT Triton 24.5%
(21.0–28.3%)
78.5%
(70.1–85.4%)
83.8%
(78.2–88.3%)
18.6%
(17.1–20.3%)

OCTA, optical coherence tomography angiography.