Values for sensitivity, specificity, predictive values and likelihood ratio for detecting obstructions (FEV1/FVC < 0.7 by spirometry) for two cutoff points of the FEV1/FEV6 ratio measured with the Vitalograph COPD-6

FEV1/FEV6 Sensitivity% Specificity % PPV% NPV% + LR −LR
<0.7 88% 98 89 97 42 0.16
<0.72 84 97 84 97 28 0.16