Apparent diffusion coefficient values measured for each acquisition for varying PVP concentrations

PVP concentration
(%)
Apparent Diffusion Coefficient (10−3 mm2/s) Coefficient of variation (%)
ss-EPI MUSE two shots MUSE three shots
Acceleration Factor Acceleration Factor
1 1.5 2 1 1.5 2
0 2.40 2.39 2.43 2.41 2.42 2.43 2.41 1.2
10 1.74 1.75 1.79 1.79 1.79 1.81 1.82 1.5
14 1.58 1.48 1.48 1.48 1.49 1.50 1.51 1.9
18 1.37 1.29 1.31 1.31 1.31 1.32 1.33 1.8
25 1.19 1.09 1.10 1.11 1.11 1.12 1.13 1.6
40 0.67 0.60 0.70 0.69 0.69 0.66 0.67 1.6