Skip to main content
. 2021 Mar 22;1(3):e85. doi: 10.1002/cpz1.85

Table 2.

Summary of Common Problems Encountered and Their Solutions

Problem Solution
Basic Protocol
Sample clogged on S‐Trap™ Centrifuge the sample thoroughly before pipetting, and be sure to shear DNA with sonication or with the addition of MgCl2 and Benzonase.
Contaminants in MS data, such as polyethylene glycol or lipids, or high backpressure during sample loading Perform additional wash steps on S‐trap™ at step 14. For removal of lipids, add three washes of 50:50 (v/v) methanol/chloroform before standard S‐trap™ buffer washes.
Percentage of peptides identified with at least one missed cleavage is >15% Increase the concentration of trypsin if digesting a large amount of protein.
Alternate Protocol 1
Sample contaminated with detergents or polymers, as evidenced by poor LCMS data Use detergent‐free glassware; be sure to perform SP2 cleanup, as C18 cleanup protocols do not remove PEG.
Large number of missed cleavages in database search results Check that the digestion solution pH is ∼8 after adding trypsin to the protein, and adjust with small amounts of concentrated acid or base as necessary.
Alternate Protocol 2
Low peptide recovery

During particle preparation, let particles warm to room temperature under constant shaking on an orbital shaker for ≥30 min. Sufficiently suspend particles by shaking and vortexing to yield a homogenous suspension. Particle preparation should be made fresh every 6 months.

Ensure that math was performed correctly during peptide quantification assay. Alternatively, repeat peptide quantification assay.

To maximize recovery, use 0.2‐0.4 µl of 50 µg/µl particle suspension per microgram of peptide to be cleaned (i.e., a particle to peptide ratio of 10:1‐20:1).

PEG or other contaminants in MS data

Use detergent‐free glassware.

Make fresh elution solution.

Support Protocols 1 and 2
No protein in sample, or low protein concentration after lysis Use ionic detergent compatibility reagent included with the Pierce kit.
Variable numbers in duplicate values Be sure to allow the reagents to come to room temperature. Consider checking pipet technique.
No protein in sample, or low protein concentration after tissue lysis Be sure to use ionic detergent compatibility reagent included in the Pierce kit if there are detergents in the sample.
Too much protein for standard range Dilute sample further with lysis buffer before attempting protein quantification again.
Support Protocol 3
Variable peptide amounts in fractions Consider changing acetonitrile percentages to optimize spread of peptides.
No peptides in any fractions Check the saved flowthrough and wash tubes for non‐retained peptides.