Table 4.
Agreement of IOP and CDR measurement by an ophthalmologist at the base hospital and technicians at the vision center
| Comparison of IOP measurement by an ophthalmologist at the base hospital and technicians at the vision center | ||||||||
|---|---|---|---|---|---|---|---|---|
| Eye | Base hospital | Vision center | p | Intraclass correlation coefficient (ICC) | ||||
| n | Mean (SD) | Range | n | Mean (SD) | Range | |||
| Right eye | 274 | 18.42±9.0 | 2-62 | 274 | 17.36±7.5 | 10-60 | 0.134 | 0.82 (0.76-0.85) |
| Left eye | 277 | 18.06±8.3 | 5-58 | 277 | 16.89±6.9 | 10-52 | 0.192 | 0.74 (0.67-0.79) |
| Both eyes | 551 | 18.25±8.7 | 2-62 | 551 | 17.12±7.2 | 10-60 | 0.163 | 0.78 (0.74-0.81) |
| Comparison of CDR measurement by an ophthalmologist at the base hospital and technicians at the vision center | ||||||||
| Eye | Base Hospital | Vision Center | p | Intraclass correlation coefficient (ICC) | ||||
| n | Mean (SD) | Range | n | Mean (SD) | Range | |||
| Right eye | 212 | 0.63±0.17 | 0.2-0.95 | 212 | 0.70±0.13 | 0.4-0.9 | 0.752 | 0.89 (0.85-0.91) |
| Left eye | 212 | 0.63±0.18 | 0.2-0.95 | 212 | 0.70±0.13 | 0.4-0.9 | 0.388 | 0.91 (0.89-0.93) |
| Both eyes | 424 | 0.63±0.18 | 0.2-0.95 | 424 | 0.70±0.13 | 0.4-0.9 | 0.57 | 0.90 (0.88-0.92) |
Values were presented with ICC and 95% confidence interval. The Intraclass Correlation Coefficient (ICC) shows good reliability (ICC >0.75) that exists between the measurement of IOP and CDR at the base hospital and vision center.