Table 3.

Validation and performance evaluation of RealDetect™ Covid-19 RT-PCR kit.


Singleplex method
Multiplex method
Result Test by RealDetect Covid-19 RT-PCR kit Test by xABT Multiple Real-time PCR kit Test by RealDetect Covid-19 RT-PCR kit Test by A* Star Fortitute kit Test by RealDetect Covid-19 RT-PCR kit Test by xABT Multiple Real-time PCR kit Test by RealDetect Covid-19 RT-PCR kit Test by A* Star Fortitute kit
Samples tested positive,
N = 120
Strong
(Ct ≤26)
N=38
18 18 20 20 18 18 20 20
Moderate
(Ct >26 to ≤32)
N = 53
27 27 26 26 27 27 26 26
Low
(Ct>32 to ≤38)
N= 29
14 15 14 14 15 15 14 14
Samples tested negative, N= 120 61 60 60 60 60 60 60 60
Total sample, n 120 120 120 120 120 120 120 120
Sensitivity,% (95 % CI) 98.33 (91.06–99.96) 100 (93.98–100) 100 (93.98–100) 100 (93.98–100)
Specificity, % (95 % CI) 100 (94.04–100) 100 (93.98–100) 100 (93.98–100) 100 (93.98–100)
Positive predictive value (95 %CI) 100 (93.94–100) 100 (93.98–100) 100 (93.98–100) 100 (93.98–100)
Negative predictive value (95 %CI) 98.36 (91.2–99.96) 100 (93.98–100) 100 (93.98–100) 100 (93.98–100)
Accuracy, % (95 % CI) 99.17 (95.43–99.85) 100 (96.9–100) 100 (96.9–100) 100 (96.9–100)
Kappa value (95 % CI) 0.98 (0.8044–1.162) 1 (0.8211–1.179) 1 (0.8211–1.179) 1 (0.8211–1.179)

Abbreviations: CI, Confidence Interval.