Table 3.
Validation and performance evaluation of RealDetect™ Covid-19 RT-PCR kit.
Singleplex method |
Multiplex method |
||||||||
---|---|---|---|---|---|---|---|---|---|
Result | Test by RealDetect Covid-19 RT-PCR kit | Test by xABT Multiple Real-time PCR kit | Test by RealDetect Covid-19 RT-PCR kit | Test by A* Star Fortitute kit | Test by RealDetect Covid-19 RT-PCR kit | Test by xABT Multiple Real-time PCR kit | Test by RealDetect Covid-19 RT-PCR kit | Test by A* Star Fortitute kit | |
Samples tested positive, N = 120 |
Strong (Ct ≤26) N=38 |
18 | 18 | 20 | 20 | 18 | 18 | 20 | 20 |
Moderate (Ct >26 to ≤32) N = 53 |
27 | 27 | 26 | 26 | 27 | 27 | 26 | 26 | |
Low (Ct>32 to ≤38) N= 29 |
14 | 15 | 14 | 14 | 15 | 15 | 14 | 14 | |
Samples tested negative, N= 120 | 61 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | |
Total sample, n | 120 | 120 | 120 | 120 | 120 | 120 | 120 | 120 | |
Sensitivity,% (95 % CI) | 98.33 (91.06–99.96) | 100 (93.98–100) | 100 (93.98–100) | 100 (93.98–100) | |||||
Specificity, % (95 % CI) | 100 (94.04–100) | 100 (93.98–100) | 100 (93.98–100) | 100 (93.98–100) | |||||
Positive predictive value (95 %CI) | 100 (93.94–100) | 100 (93.98–100) | 100 (93.98–100) | 100 (93.98–100) | |||||
Negative predictive value (95 %CI) | 98.36 (91.2–99.96) | 100 (93.98–100) | 100 (93.98–100) | 100 (93.98–100) | |||||
Accuracy, % (95 % CI) | 99.17 (95.43–99.85) | 100 (96.9–100) | 100 (96.9–100) | 100 (96.9–100) | |||||
Kappa value (95 % CI) | 0.98 (0.8044–1.162) | 1 (0.8211–1.179) | 1 (0.8211–1.179) | 1 (0.8211–1.179) |
Abbreviations: CI, Confidence Interval.