Skip to main content
. 2021 Apr 1;138:104817. doi: 10.1016/j.jcv.2021.104817

Table 4.

Diagnostic performance of the eazyplex® SARS-CoV-2 assay for prospective analysis of nasal and oropharyngeal swabs in comparison to NeuMoDx™ RT-PCR.

NeuMoDx™ eazyplex®
PPA, % (CI) NPA, % (CI) PPV, % NPV, % (CI)
Positive Negative Total
All Ct values Positive 16 8 24 66.7 (44.7−84.4) 100 (98.8−100) 100 93 (88.3−95.9)
Negative 0 106 106
Total 16 114 130
Ct values ≤30a Positive 16 3 19 84.2 (60.4−96.6) 100 (96.6−100) 100 97.2 (92.6−99)
Negative 0 106 106
Total 16 109 125
Ct values ≤28a Positive 13 0 13 100 (76.8−100) 100 (96.6−100) 100 100
Negative 0 106 106
Total 13 106 119

PPA, positive percent agreement; NPA, negative percent agreement; CI, 95 % confidence interval; PPV, positive predictive value; NPV, negative predictive value.

a

for N or/and Nsp2 gene.