Table 3.

Detailed sequence parameters of compositional MRI sequences

Imaging parameter WASSR gagCEST T1 map T2 map
FOV (mm) 160 × 160 160 × 160 160 × 160 160 × 160
Slice thickness (mm) 5 5 7 3 mm
TE (ms) 3.5 3.5 11 13.8/27.6, 41.4/55.2/69
TR (ms) 7.2 7.2 6000 1000
TI (ms) 25/50/100/500/1000/2000
Resolution (mm/pixel) 0.6 × 0.6 0.6 × 0.6 0.6 × 0.6 0.4 × 0.4
Flip angle (°) 15 15 180 180
Pulsed CEST saturation module
 Frequency range (ppm – ppm) − 1 to 1 − 3 to 3
 Number of Dynamic Scans 21 + 1 25 + 1 reference image
 Number of saturation pulses 1 8
 Pulse Duration tp (ms) 54 300 (100, 200)
 Interpulse Duration (ms) 300
 B1 amplitude (µT) 0.2 0.8 (0.6, 1.0)

In healthy volunteers, pulse duration tp and B1 amplitude were evaluated at 100, 200, and 300 ms and at 0.6, 0.8, and 1.0 µT, respectively, while in patients, the following parameter settings were used: 300 ms and 0.8 µT

WASSR water saturation, gagCEST glycosaminoglycan chemical exchange saturation transfer imaging, FOV field of view, TE echo time, TR repetition time, TI inversion time