Skip to main content
. 2021 Apr 3;342:129899. doi: 10.1016/j.snb.2021.129899

Table 1.

Comparison of CHA-LFIA and RT-PCR for SARS-CoV-2 RNA reference material detection.

Copies/mL RT-PCR
CHA-LFIA fluorescence values
Result Ct Value
Orf1ab
N
Orf1ab N Test 1 Test 2 Test 3 Test 1 Test 2 Test 3
500,000 Positive 18.25 17.33 11,960 8950 8997 9463 8215 10,448
100,000 Positive 20.43 18.38 5669 4430 5401 4688 5805 3752
20,000 Positive 21.68 19.99 2319 2800 1921 1707 1695 2725
4000 Positive 29.62 28.50 892 598 734 523 817 825
800 Positive 36.59 35.63 192 131 72 167 100 198
160 Negative 39.16 38.32 180 125 123 102 99 186
0 Negative 0 0 95 69 79 93 133 140