Skip to main content
. 2020 Apr 22;35(2):448–454. doi: 10.1038/s41433-020-0868-2

Table 1.

Mean, standard deviation and range of measurements and of differences between measurements obtained from GAT performed by nurses and ophthalmologists and ORA performed by technicians.

Mean ± sd measurements (range)
IOP GAT nurse (mmHg) 15.0 ± 4.0 (6.0 to 32.0)
IOP GAT ophthalmologist (mmHg) 14.6 ± 4.3 (5.0 to 33.5)
ORA IOPg1 (mmHg) 13.8 ± 4.7 (3.8 to 31.5)
ORA IOPg2 (mmHg) 12.8 ± 4.7 (3.4 to 28.1)
ORA IOPcc1 (mmHg) 15.1 ± 4.2 (7.1 to 32.3)
ORA IOPcc2 (mmHg) 13.8 ± 4.2 (7.9 to 28.9)
ORA WS1 8.4 ± 1.0 (3.4 to 9.6)
ORA WS2 8.4 ± 1.0 (4.1 to 9.6)
ORA CH1 (mmHg) 9.9 ± 1.3 (6.5 to 13.3)
ORA CH2 (mmHg) 10.2 ± 1.4 (7.3 to 15.1)
CCT (μm) 540 ± 44 (445 to 658)
IOP GAT nurse minus IOP GAT ophthalmologist (mmHg) 0.4 ± 2.9 (−12 to 8)
IOPg minus IOP GAT ophthalmologist (mmHg) −0.8 ± 2.5 (−7.5 to 5.1)
IOPg minus IOP GAT nurse (mmHg) −1.2 ± 3.2 (−13.5 to 7.9)
IOPg1 minus IOPg2 (mmHg) 0.9 ± 1.3 (−2.9 to 4.9)
CH1 minus CH2 (mmHg) −0.2 ± 0.6 (−3 to 1.9)
IOPcc1 minus IOPcc2 (mmHg) 1.0 ± 1.4 (−2.4 to 5.5)