Table 2. Validity between Veloflex and CROM to measure the cervical range of motion.

Cervical movement Mean CROM (SD) Mean veloflex (SD) Pearson coefficient LoA− (%) LoA+ (%) Mean difference (%) SD difference (%)
Flexion 61.17° (8.60°) 61.35° (8.78°) 0.993 −1.83° (2.99%) 2.18° (3.56%) 0.17° (0.27%) 1.02° (1.66%)
Extension 61.22° (10.70°) 61.38° (10.7°) 0.998 −1.26° (2.05%) 1.60° (2.61%) 0.17° (0.27%) 0.73° (1.19%)
Right side bending 40.40° (11.01°) 40.56° (11.20°) 0.995 −2.02° (5%) 2.34° (5.79%) 0.16° (0.39%) 1.11° (2.74%)
Left side bending 40.63° (8.85°) 40.46° (8.90°) 0.993 −2.30° (5.66%) 1.96° (4.82%) −0.17° (0.41%) 1.09° (2.68%)
Right rotation 74.30° (8.53°) 73.33° (7.51°) 0.974 −1.90° (2.55%) 2.15° (2.89%) −0.96° (1.29%) 2.10° (2.82%)
Left rotation 75.49° (7.57°) 74.24° (7.24°) 0.960 −1.15° (1.52%) 1.43° (1.89%) −1.24° (1.64%) 2.19° (2.90%)

Note:

SD, standard deviation; LoA, limit of agreement