Perform the clinical accuracy protocol described in ANSI/AAMI/ISO 81060‐2 |
Establishes the BP device's clinical accuracy on stable, resting patients |
Download standardized set of noise files from the AAMI‐hosted noise library |
Standardized noise files assure that the performance of different BP devices is evaluated equally. |
Preload a noise file into the noise playback device |
Set up the noise playback device |
Connect the BP simulator equipment to the device under test as depicted in the 81060‐4 draft |
Assure a standardized setup |
Configure the BP simulator to one of the predetermined BP values |
Assure a standardized set of BP simulations |
Perform and record the device readings |
Establish baseline performance with no noise present |
Connect the BP simulator and noise playback device to the device under test as depicted in the 81060‐4 draft |
Prepare to perform the same set of BP determinations with added artifact noise. Note: the BP simulation and noise playback functionality may be integrated into a single piece of test equipment. |
Configure both the BP simulator and the noise playback simulator to one of the predetermined combinations |
Assure a standardized set of BP + noise simulations |
Perform and record the device readings |
Measure the performance with noise present |
Gather a set of device readings for each predetermined combination listed in the 81060‐4 draft |
Assure a consistent set of noise artifact challenges. This set of challenges is intended to represent a realistic subset of the artifact noise spectrum. |
Compare and analyze the groups of device readings according to the instructions and specifications detailed in 81060‐4 draft. |
The results analysis will indicate the impact of artifact noise on the BP device reading accuracy and device behavior |